A rapid method for measuring intraocular lens power in vitro with a focimeter
Por:
García-Domene MC, Díez-Ajenjo MA, Peris-Martínez C, Navea A and Artigas JM
Publicada:
1 nov 2015
Resumen:
In this paper we describe a new method for measuring the intraocular lens (IOL) power using a focimeter, a negative ophthalmic lens and a saline solution (0.9% NaCl). To test this we measured the power of 58 different IOLs and we compared them with the power stated by the manufacturer. Despite the limitations, the results show a good correlation. (C) 2015 Elsevier Ltd. All rights reserved.
Filiaciones:
:
Universitat de València, C/Dr. Moliner, 50, 46100 Burjassot, Valencia, Spain.
FISABIO-Oftalmología Médica, Bifurcación Pío Baroja-General Aviles, s/n, 46015 Valencia, Spain
:
Universitat de València, C/Dr. Moliner, 50, 46100 Burjassot, Valencia, Spain
FISABIO-Oftalmología Médica, Bifurcación Pío Baroja-General Aviles, s/n, 46015 Valencia, Spain
:
Universitat de València, C/Dr. Moliner, 50, 46100 Burjassot, Valencia, Spain
|