A rapid method for measuring intraocular lens power in vitro with a focimeter


Por: García-Domene MC, Díez-Ajenjo MA, Peris-Martínez C, Navea A and Artigas JM

Publicada: 1 nov 2015
Resumen:
In this paper we describe a new method for measuring the intraocular lens (IOL) power using a focimeter, a negative ophthalmic lens and a saline solution (0.9% NaCl). To test this we measured the power of 58 different IOLs and we compared them with the power stated by the manufacturer. Despite the limitations, the results show a good correlation. (C) 2015 Elsevier Ltd. All rights reserved.

Filiaciones:
:
 Universitat de València, C/Dr. Moliner, 50, 46100 Burjassot, Valencia, Spain.

 FISABIO-Oftalmología Médica, Bifurcación Pío Baroja-General Aviles, s/n, 46015 Valencia, Spain

:
 Universitat de València, C/Dr. Moliner, 50, 46100 Burjassot, Valencia, Spain

 FISABIO-Oftalmología Médica, Bifurcación Pío Baroja-General Aviles, s/n, 46015 Valencia, Spain

:
 Universitat de València, C/Dr. Moliner, 50, 46100 Burjassot, Valencia, Spain
ISSN: 00144835





EXPERIMENTAL EYE RESEARCH
Editorial
ACADEMIC PRESS LTD- ELSEVIER SCIENCE LTD, 24-28 OVAL RD, LONDON NW1 7DX, ENGLAND, Estados Unidos America
Tipo de documento: Article
Volumen: 140 Número:
Páginas: 190-192
WOS Id: 000364259700022
ID de PubMed: 26386149

MÉTRICAS